标题:Near infrared spectroscopy combined with multivariate analysis for monitoring the ethanol precipitation process of fraction I + II + III supernatant in human albumin separation
作者:Li, Can ;Wang, Fei ;Zang, Lixuan ;Zang, Hengchang ;, Manel ;Nie, Lei ;Wang, Mingyu ;Li, Lian
作者机构:[Li, Can ;Wang, Fei ;Zang, Lixuan ;Zang, Hengchang ;Zang, Hengchang ;Nie, Lei ;Li, Lian ] School of Pharmaceutical Sciences and National Glycoengineer 更多
通讯作者:Zang, Hengchang
来源:Spectrochimica Acta - Part A: Molecular and Biomolecular Spectroscopy
出版年:2017
卷:175
页码:17-23
DOI:10.1016/j.saa.2016.12.016
关键词:Ethanol precipitation process; Human albumin; Near infrared spectroscopy; Qualitative analysis; Quantitative determination
摘要:Nowadays, as a powerful process analytical tool, near infrared spectroscopy (NIRS) has been widely applied in process monitoring. In present work, NIRS combined with multivariate analysis was used to monitor the ethanol precipitation process of fraction I + II + III (FI + II + III) supernatant in human albumin (HA) separation to achieve qualitative and quantitative monitoring at the same time and assure the product's quality. First, a qualitative model was established by using principal component analysis (PCA) with 6 of 8 normal batches samples, and evaluated by the remaining 2 normal batches and 3 abnormal batches. The results showed that the first principal component (PC1) score chart could be successfully used for fault detection and diagnosis. Then, two quantitative models were built with 6 of 8 normal batches to determine the content of the total protein (TP) and HA separately by using partial least squares regression (PLS-R) strategy, and the models were validated by 2 remaining normal batches. The determination coefficient of validation (Rp 2), root mean square error of cross validation (RMSECV), root mean square error of prediction (RMSEP) and ratio of performance deviation (RPD) were 0.975, 0.501 g/L, 0.465 g/L and 5.57 for TP, and 0.969, 0.530 g/L, 0.341 g/L and 5.47 for HA, respectively. The results showed that the established models could give a rapid and accurate measurement of the content of TP and HA. The results of this study indicated that NIRS is an effective tool and could be successfully used for qualitative and quantitative monitoring the ethanol precipitation process of FI + II + III supernatant simultaneously. This research has significant reference value for assuring the quality and improving the recovery ratio of HA in industrialization scale by using NIRS. © 2016 Elsevier B.V.
收录类别:EI;SCOPUS
Scopus被引频次:2
资源类型:期刊论文
原文链接:https://www.scopus.com/inward/record.uri?eid=2-s2.0-85006826434&doi=10.1016%2fj.saa.2016.12.016&partnerID=40&md5=a96187096074a09e05401fc3bedb075a
TOP