标题:Magnetic field simulation and shimming analysis of 3.0T superconducting MRI system
作者:Yue, Z. K.; Liu, Z. Z.; Tang, G. S.; Zhang, X. C.; Duan, L. J.; Liu, W. C.
通讯作者:Yue, ZK
作者机构:[Yue, Z. K.; Liu, Z. Z.; Tang, G. S.; Zhang, X. C.; Duan, L. J.; Liu, W. C.] Shandong Univ, Sch Elect Engn, Jinan 250061, Shandong, Peoples R China.
会议名称:Asia Conference on Energy and Environment Engineering (ACEEE)
会议日期:JAN 19-21, 2018
来源:2018 ASIA CONFERENCE ON ENERGY AND ENVIRONMENT ENGINEERING (ACEEE 18)
出版年:2018
卷:133
期:1
DOI:10.1088/1755-1315/133/1/012003
摘要:3.0T superconducting magnetic resonance imaging (MRI) system has become the mainstream of modern clinical MRI system because of its high field intensity and high degree of uniformity and stability. It has broad prospects in scientific research and other fields. We analyze the principle of magnet designing in this paper. We also perform the magnetic field simulation and shimming analysis of the first 3.0T/850 superconducting MRI system in the world using the Ansoft Maxwell simulation software. We guide the production and optimization of the prototype based on the results of simulation analysis. Thus the magnetic field strength, magnetic field uniformity and magnetic field stability of the prototype is guided to achieve the expected target.
收录类别:CPCI-S;EI;SCOPUS